Patient radiation exposure in a modern, large-volume, pediatric cardiac catheterization laboratory
Andrew C Glatz1, Akash Patel, Xiaowei Zhu
1Division of Cardiology, The Children's Hospital of Philadelphia, 6th Floor, Main Building, 34th Street and Civic Center Blvd., Philadelphia, PA, 19104, USA, glatz@email.chop.edu.
Insights
Pediatric cardiac catheterization involves substantial radiation exposure, with a median effective dose of 6.2 mSv. Radiation dose varies significantly by procedure type and patient factors, necessitating ongoing dose reduction strategies.
Area of Science:
- Cardiology
- Radiology
- Pediatric Medicine
Background:
- Pediatric cardiac catheterization can involve significant radiation exposure.
- Published estimates of radiation dose vary widely.
- Optimizing radiation safety in pediatric procedures is crucial.
Purpose of the Study:
- To report patient radiation dose in diagnostic and interventional pediatric cardiac catheterization.
- To analyze radiation exposure across various procedures in a high-volume laboratory.
- To identify predictors of radiation dose in pediatric patients.
Main Methods:
- Retrospective review of 2,265 diagnostic and interventional cases (excluding electrophysiology).
- Radiation dose quantified using air kerma dose (mGy) and dose-area product (DAP; μGy m²).
- Effective dose (mSv) calculated from DAP using Monte Carlo methods.
Main Results:
- Median effective dose was 6.2 mSv (IQR 2.7–14.1 mSv).
- Interventional procedures had higher effective doses than diagnostic ones (p < 0.001).
- Patient weight, age, fluoroscopy time, and digital acquisition use predicted exposure (R² = 0.59–0.64).
Conclusions:
- Median effective radiation dose in pediatric cardiac catheterization is 6.2 mSv.
- Radiation exposure is highly variable, influenced by patient and procedural factors.
- Radiation monitoring and dose reduction strategies are essential in pediatric catheterization labs.
Abstract:
Radiation exposure from pediatric cardiac catheterization may be substantial, although published estimates vary. We sought to report patient radiation dose across a range of diagnostic and interventional cases in a modern, high-volume pediatric catheterization laboratory. We retrospectively reviewed diagnostic and interventional cases performed in our pediatric catheterization laboratory from 1 April 2009 to 30 September 2011 for which radiation usage data were available as reported by the Artis Zee(®) (Siemens Medical Solutions) system. Electrophysiology cases were excluded. Radiation dose was quantified as air kerma dose (mGy) and dose-area product (DAP; μGy m(2)). The DAP was converted to an effective dose millisievert (mSv) using the Monte Carlo method. Radiation usage data were available from 2,265 diagnostic and interventional cases with an overall median air kerma dose of 135 mGy [interquartile range (IQR) 59-433], median DAP of 760 μGy m(2) (IQR 281-2,810), of which 75 % (IQR 59-90 %) was derived from fluoroscopy, and median effective dose of 6.2 mSv (IQR 2.7-14.1). Air kerma dose from a single camera >2,000 mGy occurred in 1.8 % of cases. Significant differences in all measures of radiation exposure existed based on procedural and interventional types (p = 0.0001), with interventional cases associated with the highest effective dose after adjusting for patient weight category (p < 0.001). Patient weight, age, fluoroscopy time, and proportional use of digital acquisition were independent predictors of exposure (p ≤ 0.001; R (2) = 0.59-0.64). In a modern, large-volume pediatric catheterization laboratory, the median effective dose is 6.2 mSv with a wide range of exposure based on patient- and procedure-specific factors. Radiation monitoring is an important component of a pediatric laboratory and further dose reduction strategies are warranted.
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiac Catheterization IV: Nursing Management