Related Experiment Video
Updated: May 3, 2026

Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
Universal intravenous access cleaning device fails to sterilize stopcocks
Julie L Holroyd1, David A Paulus, Kenneth H Rand
1From the Departments of Anesthesiology and Pathology, Immunology, and Laboratory Medicine, University of Florida College of Medicine, Gainesville, Florida; Deceased.
Background:
Contamination of a central venous catheter may occur through use of conventional open-lumen stopcock devices (COLDs), or disinfectable, needleless, closed connectors (DNCCs). We investigated the effectiveness of a new universal IV access cleaning device (Site-Scrub) compared with 70% isopropyl alcohol prep pads for sanitizing COLDs or DNCCs inoculated with common catheter-associated pathogens.
Methods:
Site-Scrub was compared with 70% alcohol prep pads for sanitizing contaminated female Luer lock COLD or DNCC filled with sterile saline or propofol and 2 common bacterial central venous catheter contaminants (Staphylococcus epidermidis or Pseudomonas aeruginosa). Devices were contaminated using a glove touch (COLD and DNCC) or syringe tip (COLD). The primary end point of the study was colony-forming units (CFU) after 24 hours.
Results:
The use of glove touch contamination, the contaminants, S epidermidis and P aeruginosa, produced CFU in saline-filled COLDs treated with the Site-Scrub, but not in those treated with alcohol pads (P < 0.001). Similar results were observed with propofol-filled COLDs (P < 0.001). For DNCCs filled with saline or propofol, both alcohol and Site-Scrub effectively reduced CFU growth compared with contaminated controls (P < 0.001). When COLDs were contaminated by treated syringe tips, there was no significant evidence of reduction in CFU growth by using either alcohol pads or Site-Scrub compared with contaminated controls.
Conclusions:
These data suggest that when the inner surface of the COLD is contaminated, both alcohol pads and Site-Scrub were not significantly effective in decontaminating the COLD. When the COLD rim is contaminated, however, alcohol pads outperform Site-Scrub. DNCCs were uniformly decontaminated with either treatment. Future work should focus on better access systems because current COLDs are difficult to decontaminate.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hemodialysis II: Procedure and Complications
Tracheostomy Suctioning II: Procedure
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...