Related Experiment Video
Updated: May 3, 2026

Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
Universal intravenous access cleaning device fails to sterilize stopcocks
Julie L Holroyd1, David A Paulus, Kenneth H Rand
1From the Departments of Anesthesiology and Pathology, Immunology, and Laboratory Medicine, University of Florida College of Medicine, Gainesville, Florida; Deceased.
A new Site-Scrub device showed limited effectiveness in decontaminating conventional open-lumen stopcock devices (COLDs) compared to alcohol pads. Disinfectable, needleless, closed connectors (DNCCs) were effectively decontaminated by both methods.
Area of Science:
- Infection control
- Medical device decontamination
- Microbiology
Background:
- Central venous catheter (CVC) contamination poses a risk.
- Conventional open-lumen stopcock devices (COLDs) and disinfectable, needleless, closed connectors (DNCCs) are common access points.
- Pathogen inoculation of these devices can lead to catheter-associated infections.
Purpose of the Study:
- To compare the efficacy of a new universal IV access cleaning device, Site-Scrub, against 70% isopropyl alcohol prep pads.
- To evaluate the sanitization of contaminated COLDs and DNCCs using common CVC pathogens.
- To determine the effectiveness of different decontamination methods for various contamination scenarios.
Main Methods:
- Site-Scrub and alcohol prep pads were used to sanitize contaminated COLDs and DNCCs.
- Devices were filled with sterile saline or propofol and inoculated with Staphylococcus epidermidis or Pseudomonas aeruginosa.
- Contamination was introduced via glove touch or syringe tip, with colony-forming units (CFU) assessed after 24 hours.
Main Results:
- Site-Scrub showed limited efficacy on inner surface contamination of COLDs, allowing CFU growth, unlike alcohol pads.
- Alcohol pads were more effective than Site-Scrub when the rim of COLDs was contaminated.
- Both Site-Scrub and alcohol pads effectively reduced CFU growth in DNCCs, regardless of fill solution.
Conclusions:
- Current conventional open-lumen stopcock devices (COLDs) present challenges for effective decontamination.
- Disinfectable, needleless, closed connectors (DNCCs) demonstrate superior decontamination outcomes with both tested methods.
- Further research into improved IV access systems is warranted to enhance CVC safety.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hemodialysis II: Procedure and Complications
Tracheostomy Suctioning II: Procedure
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...