Related Experiment Video
Updated: May 3, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Evaluation of the Nova StatStrip glucometer in a pediatric hospital setting
M G Lockyer1, K Fu1, R M Edwards2
1Baylor College of Medicine, 1 Baylor Plaza MC 315, Houston, TX, 77030, USA.
Context:
Routine point-of-care (POC) glucose monitoring in the pediatric setting has become increasingly important, both for assessing hypoglycemia as well as hyperglycemia. A reliable and precise system is required to monitor pediatric patients.
Objectives:
The aim of this study was to evaluate the Nova Biomedical StatStrip POC glucometer against the Roche ACCU-CHEK Inform in lieu of our currently used LifeScanSureStepFlexx POC glucose analyzer.
Design And Methods:
Intra-assay and inter-assay precision, linearity, correlation and interference studies were performed as per the NCCLS criteria. An analysis of 37 pediatric samples across the linearity ranges of all the meters was used to assess concordance between the systems.
Results:
The Nova StatStrip glucometer demonstrated an excellent coefficient of variation (<5%) for glucose across the entire analytical measurement range. The Nova StatStrip also had good concordance with the central laboratory (Bland-Altman plots r(2)=0.01), while Roche Inform had poorer correlation (Bland-Altman plots r(2)=0.46). We also evaluated the effect of hematocrit (20-60%) and maltose (100-500mg/dL) on the Nova StatStrip analyzer and demonstrated that there is little to no interference by either.
Conclusions:
The Nova StatStrip system gave the best performance with acceptable imprecision, good correlation, and minimal to no interference from hematocrit levels or maltose. The Nova StatStrip is a satisfactory replacement for our POC glucometer system and, additionally, provides results in less time (just 6 s) utilizing a lower amount of blood with the advantage of being immediately interfaced to our laboratory information systems.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...