Clinically relevant quality assurance (QA) for prostate RapidArc plans: gamma maps and DVH-based evaluation
M Cozzolino1, C Oliviero1, G Califano1
1Department of Radiation Oncology, IRCCS CROB, 1 Padre Pio Street, 85028 Rionero in Vulture, PZ, Italy.
Quality assurance tests for RapidArc prostate radiotherapy were evaluated. Per-structure gamma analysis, not whole-grid, better detects dose deviations, improving patient safety.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Quality Assurance
Background:
- RapidArc radiotherapy offers precise dose delivery for prostate cancer.
- Independent verification of treatment plans is crucial for patient safety.
- Evaluating quality assurance (QA) methods ensures treatment accuracy.
Purpose of the Study:
- To clinically evaluate quality assurance (QA) tests for RapidArc prostate patients.
- To compare the effectiveness of whole-grid versus per-structure gamma analysis.
- To assess the correlation between gamma passing rates and dose-volume histogram (DVH) discrepancies.
Main Methods:
- Verified 26 RapidArc prostate plans using the COMPASS system for independent dose reconstruction.
- Performed gamma analysis with 3%-3mm and 2%-2mm criteria on whole-grid and per-structure bases.
- Conducted DVH-based analysis for target and organs-at-risk (OAR), correlating with gamma passing rates using Pearson's test.
Main Results:
- No significant DVH deviations were observed for target and OAR.
- A weak negative correlation was found between gamma passing rates and dosimetric deviations.
- The whole-grid gamma method lacked sensitivity for deviations >5%, while per-structure gamma showed better sensitivity and specificity.
Conclusions:
- Per-structure gamma analysis, particularly with 2%-2mm criteria, is recommended for improved sensitivity and specificity in RapidArc prostate QA.
- DVH-based analysis is encouraged when feasible due to poor correlation between standard gamma passing rates and DVH goals.
- Structure-specific gamma methods are strongly suggested for robust radiotherapy quality assurance.