Related Experiment Video
Updated: May 3, 2026

Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Design process of cementless femoral stem using a nonlinear three dimensional finite element analysis
Mohd Yusof Baharuddin, Sh-Hussain Salleh1, Ahmad Hafiz Zulkifly
1Centre for Biomedical Engineering Transportation Research Alliance, Universiti Teknologi Malaysia, Skudai, Johor, Malaysia. hussain@fke.utm.my.
Background:
Minimal available information concerning hip morphology is the motivation for several researchers to study the difference between Asian and Western populations. Current use of a universal hip stem of variable size is not the best option for all femur types. This present study proposed a new design process of the cementless femoral stem using a three dimensional model which provided more information and accurate analysis compared to conventional methods.
Methods:
This complete design cycle began with morphological analysis, followed by femoral stem design, fit and fill analysis, and nonlinear finite element analysis (FEA). Various femur parameters for periosteal and endosteal canal diameters are measured from the osteotomy level to 150 mm below to determine the isthmus position.
Results:
The results showed better total fit (53.7%) and fill (76.7%) canal, with more load distributed proximally to prevent stress shielding at calcar region. The stem demonstrated lower displacement and micromotion (less than 40 μm) promoting osseointegration between the stem-bone and providing primary fixation stability.
Conclusion:
This new design process could be used as a preclinical assessment tool and will shorten the design cycle by identifying the major steps which must be taken while designing the femoral stem.