A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Role of defibrillation threshold testing during implantable cardioverter-defibrillator placement: data from the
Yoav Arnson1, Mahmoud Suleiman2, Michael Glikson3
1Department of Cardiology, Meir Medical Center, Kfar Saba, Israel.
Background:
Defibrillation threshold (DFT) testing during placement of an implantable cardioverter-defibrillator (ICD) has been considered mandatory. Accumulating data suggest a more limited role for DFT.
Objective:
The purpose of this study was to compare the outcome of ICD recipients who underwent DFT testing compared with those who did not.
Methods:
In this prospective cohort analysis of patients who received an ICD between July 2010 and March 2013, we compared patients who underwent DFT testing and those who did not. Primary end-points were death and malignant ventricular arrhythmias. Secondary end-points included the composite end-points and inappropriate ICD discharges.
Results:
Of the 3596 patients in the registry, 614 patients (17%) underwent DFT testing during ICD placement vs 2982 (83%) who did not. Variables associated with ICD testing were implantation for secondary prevention (relative risk [RR] 1.87), prior ventricular arrhythmias (RR 1.81), use of antiarrhythmic medication (RR 1.59), and sinus rhythm (RR 2.05). Factors predisposing against testing were cardiac resynchronization therapy defibrillator implantation (RR 0.56) and concomitant diuretic use (RR 0.71). ICD testing was not associated with 1-year mortality (5.3% vs 5.1%, P = .74), delivery of appropriate shocks (8.6% vs 5.6%, P = .16), combined outcomes of ventricular arrhythmias and death (12.9% vs 11.3%, P = .45), or inappropriate ICD discharges (3.9% vs 2.1%, P = .2) compared to no DFT testing.
Conclusion:
No significant differences in the incidence of mortality, malignant ventricular arrhythmias, or inappropriate ICD discharges were observed between patients who underwent DFT testing compared to those who did not. Our results may support avoiding DFT testing during ICD placement, but this requires confirmation by additional prospective studies.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy V: Interprofessional Care