Evaluation of the Di(2-ethylhexyl)phthalate released from polyvinyl chloride medical devices that contact blood
Hongyu Luo1, Guangyu Sun1, Yanping Shi1
1CFDA Jinan Quality Inspection Center for Medical Devices, 99 Tianluo Road H-T Industrial Development Zone, Jinan, 250000 China ; Shandong Provincial Key Laboratory of Biological Evaluation of Medical Device, 99 Tianluo Road H-T Industrial Development Zone, Jinan, 250000 China.
Abstract:
Extraction methods that simulate those used in the clinic are recommended for obtaining extraction solutions. For polyvinyl chloride (PVC) medical devices that have contact with human blood, an alternative medium (ethanol/water mixture) is suggested as an extraction screening vehicle to evaluate the di(2-ethylhexyl)phthalate (DEHP) released. A test comparing the extraction ability between the alternative medium and whole blood from three healthy volunteers has been conducted. An experimental method is provided outlining the chemical analysis of the DEHP released from medical devices made with polyvinyl chloride (PVC). Gas chromatography-mass spectrometry (GC-MS) in the selective ion monitoring (SIM) mode was used to analyze and quantify the extracted DEHP. The linear range of the SIM method was 0.1-200 μg/mL, and the recoveries were 89.6-101.5% and 91.0-98.9% when using the ethanol/water mixture and whole blood as the extraction media, respectively. The validated method demonstrates that it is suitable for the determination of the DEHP released from PVC medical devices that have contact with blood. The results from the determination of the DEHP released will be compared with the limits derived from toxicological data for the parenteral exposure route and certain population groups, and the results will be used in the risk assessment of medical devices.
More Related Videos
10:15The Use of the Ex Vivo Chandler Loop Apparatus to Assess the Biocompatibility of Modified Polymeric Blood Conduits
Published on: August 20, 2014
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Related Concept Videos
Bioplastics
Microbial Bioremediation of Plastics