Fluoroscopy-guided snare retrieval of the celt ACD(®) metallic vascular closure device following failed deployment
Thomas J Cahill1, Kiyoshi Choji, Attila Kardos
1Department of Cardiology, Milton Keynes Hospital, NHS Foundation Trust, United Kingdom.
Abstract:
We report a case of endovascular snare retrieval of a new stainless steel vascular closure device (Celt ACD(®) , Kimal, Middlesex, UK) from the common femoral artery, following device failure after diagnostic coronary angiography. The stainless steel composition of the device aided successful fluoroscopic localization and removal.