Validity of a novel wristband tonometer for measuring central hemodynamics and augmentation index
Darren T Beck1, Jeffrey S Martin2, Wilmer W Nichols3
1Department of Kinesiology, University of Rhode Island, Kingston, Rhode Island; darrentbeck@mail.uri.edu.
Background:
Central hemodynamic and augmentation indices are independent predictors of cardiovascular events and all-cause mortality that can be estimated noninvasively by pulse wave analysis. The purpose of this study was to assess the reliability and validity of a newly engineered wristband tonometer for acquiring radial artery pressure waveforms.
Methods:
Radial artery pulse pressure waveforms were evaluated with an established pencil-type and a novel wristband tonometer in 31 participants (aged 30.2±9.5 years) resting in a supine position. Pulse wave analysis was executed using the same validated generalized transfer function (SphygmoCor) for both tonometers.
Results:
A significant difference in time to data acquisition between tonometers was observed (-70.2±147.7 s; P < 0.05; wristband faster). The wristband tonometer had significantly lower within-subject coefficients of variation (CV) compared with the pencil-type tonometer in aortic pulse wave height (-2.67% ± 5.51%; P < 0.05) and time to reflection (-2.26% ± 6.16%; P < 0.01). No other differences in CV were observed. Slight but statistically significant mean differences between tonometers were observed in aortic systolic blood pressure (ASBP; 0.43±1.08 mm Hg; P < 0.05; wristband lower), aortic pulse pressure (APP; 0.43±0.96 mm Hg; P < 0.05; wristband lower), and round-trip travel time of the reflected pressure wave (Δtp; 3.58±12.86 ms; P < 0.05; wristband higher). However, ASBP, APP, and Δtp measurements were highly correlated (r = 0.9970, r = 0.9953, and r = 0.8838, respectively, P <0.0001) between tonometers; within-subject and between tonometer significant mean differences were within clinical ranges.
Conclusions:
This novel, hands-free platform may be interchangeable with the commonly used pencil-type tonometer, heralding new directions in noninvasive in vivo vascular research and clinical application.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.