The temperature change in an endotracheal tube during high frequency ventilation using an artificial neonatal lung
Ken Nagaya1, Etsushi Tsuchida, Fumikatsu Nohara
1The Center for Maternity and Infant Care, Asahikawa Medical University, Hokkaido, Japan.
Objective:
There is little available data on airway humidity during high-frequency ventilation (HFV). The purpose of this study is to evaluate the temperature drop in an endotracheal tube (ETT) during HFV.
Methods:
We examined the airway temperature in a neonatal HFV system using Babylog® 8000 plus. We measured the temperature change of inspired gases in the ETT under various oscillatory frequencies and oscillatory volumes with a fixed base flow. The temperatures in the ETT during HFV were compared with the temperatures during conventional intermittent positive pressure ventilation (IPPV).
Results:
As the oscillatory frequency was increased and the oscillatory volume (VThf) decreased, the difference in temperature between the Y piece and the inlet of an artificial lung in the ETT (ETT outside of body) increased. However, as the oscillatory frequency increased, there was no difference in the ETT temperature under constant oscillatory volume. In contrast, as the oscillatory volume was decreased, the difference in temperature in the ETT was greater under constant oscillatory frequency. Moreover, the temperature drop in the ETT with HFV was lower than that in the IPPV temperature with a similar respiratory volume.
Conclusions:
The temperature change in the ETT was not dependent on the oscillatory frequency when the oscillatory volume was fixed; however, the temperature was dependent on the oscillatory volume when the oscillatory frequency was fixed.
More Related Videos
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: