Related Experiment Video
Updated: May 2, 2026

Investigation of a Blue Light LED Device to Suppress Wound Pathogens Using a Collagen-Based Synthetic Skin Model
Published on: February 24, 2026
An evaluation of the infection control potential of a UV clinical podiatry unit
Paul N Humphreys1, Chris S Davies, Simon Rout
1Hygiene and Disinfection Centre, School of Applied Sciences, University of Huddersfield, Queensgate, Huddersfield, UK. p.n.humphreys@hud.ac.uk.
Background:
Infection control is a key issue in podiatry as it is in all forms of clinical practice. Airborne contamination may be particularly important in podiatry due to the generation of particulates during treatment. Consequently, technologies that prevent contamination in podiatry settings may have a useful role. The aims of this investigation were twofold, firstly to determine the ability of a UV cabinet to protect instruments from airborne contamination and secondly to determine its ability to remove microbes from contaminated surfaces and instruments.
Method:
A UV instrument cabinet was installed in a University podiatry suite. Impact samplers and standard microbiological techniques were used to determine the nature and extent of microbial airborne contamination. Sterile filters were used to determine the ability of the UV cabinet to protect exposed surfaces. Artificially contaminated instruments were used to determine the ability of the cabinet to remove microbial contamination.
Results:
Airborne bacterial contamination was dominated by Gram positive cocci including Staphylococcus aureus. Airborne fungal levels were much lower than those observed for bacteria. The UV cabinet significantly reduced (p < 0.05) the observed levels of airborne contamination. When challenged with contaminated instruments the cabinet was able to reduce microbial levels by between 60% to 100% with more complex instruments e.g. clippers, remaining contaminated.
Conclusions:
Bacterial airborne contamination is a potential infection risk in podiatry settings due to the presence of S. aureus. The use of a UV instrument cabinet can reduce the risk of contamination by airborne microbes. The UV cabinet tested was unable to decontaminate instruments and as such could pose an infection risk if misused.
Related Concept Videos
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Hand hygiene
Hand washing...
Factors Affecting the Risk of Infection
The integrity and count of the white blood cells help the body resist pathogens and fight infection. When impaired, it reduces the body's resistance to pathogens. The acidic pH levels of the gastrointestinal, genitourinary tracts, and skin...
Diabetic Foot Ulcer
Urinary Tract Infection III: Diagnostic Studies and Interprofessional Care
Antimicrobial Effectiveness