Related Experiment Video
Updated: May 2, 2026

Clinical Microfluidic Chip Platform for the Isolation of Versatile Circulating Tumor Cells
Published on: October 13, 2023
Microfluidic, marker-free isolation of circulating tumor cells from blood samples
Nezihi Murat Karabacak1, Philipp S Spuhler1, Fabio Fachin2
11] Department of Surgery and Center for Engineering in Medicine, Massachusetts General Hospital, Boston, Massachusetts, USA. [2].
None:
The ability to isolate and analyze rare circulating tumor cells (CTCs) has the potential to further our understanding of cancer metastasis and enhance the care of cancer patients. In this protocol, we describe the procedure for isolating rare CTCs from blood samples by using tumor antigen-independent microfluidic CTC-iChip technology. The CTC-iChip uses deterministic lateral displacement, inertial focusing and magnetophoresis to sort up to 10⁷ cells/s. By using two-stage magnetophoresis and depletion antibodies against leukocytes, we achieve 3.8-log depletion of white blood cells and a 97% yield of rare cells with a sample processing rate of 8 ml of whole blood/h. The CTC-iChip is compatible with standard cytopathological and RNA-based characterization methods. This protocol describes device production, assembly, blood sample preparation, system setup and the CTC isolation process. Sorting 8 ml of blood sample requires 2 h including setup time, and chip production requires 2-5 d.