A new flow cytometric method for differential cell counting in ascitic fluid
Gert-Jan M van de Geijn1, Marc van Gent, Natasja van Pul-Bom
1Department of Clinical Chemistry (KCHL), Sint Franciscus Gasthuis, Kleiweg 500, 3045 PM, Rotterdam, The Netherlands.
Cytometry. Part B, Clinical Cytometry
|March 1, 2014
Summary
Flow cytometry offers a faster and more accurate method for counting cells in ascitic fluid compared to traditional methods. This technique is highly recommended for detecting spontaneous bacterial peritonitis (SBP) with superior sensitivity and specificity.
Area of Science:
- Clinical diagnostics
- Biomedical engineering
- Fluid analysis
Background:
- Cell counting in body fluids like ascites is challenging and time-consuming.
- Current methods involve automated counters or manual counting chambers with cytospins, possessing several limitations.
- A novel flow cytometry assay was developed for rapid ascites cell enumeration.
Purpose of the Study:
- To develop and validate a flow cytometry assay for accurate cell counting in ascitic fluid.
- To compare the performance of the flow cytometry assay against an automated cell counter (LH750) and manual cytospin analysis.
- To assess the utility of flow cytometry for detecting spontaneous bacterial peritonitis (SBP) based on polymorphonuclear neutrophil (PMN) counts.
Main Methods:
- Fifty-three ascitic fluid samples from 38 patients were analyzed.
- Flow cytometry was used to identify and quantify polymorphonuclear neutrophils (PMN), lymphocytes, eosinophils, and macrophages.
- Results were compared to a reference method combining LH750 leukocyte counts with manual differential counts from cytospins.
Main Results:
- Flow cytometry assay demonstrated excellent linearity, reproducibility, and a low detection limit, confirming its suitability for ascites cell counting.
- For detecting ≥ 250 PMN/mm³, flow cytometry achieved 100% sensitivity and 100% specificity.
- The automated LH750 showed 100% sensitivity but only 65% specificity for SBP detection.
Conclusions:
- Flow cytometry is analytically well-suited for precise cell counting in ascitic fluid.
- The flow cytometry method is significantly faster and easier to perform than current gold standards.
- Flow cytometry is recommended for rapid and accurate detection of SBP in ascitic fluid due to its high diagnostic performance.