Quantifying sampling method-induced imprecision in user-friendly lateral upper arm blood collection: Introducing
Vincent J Molenaar1, María Villalobos-Quesada2, Tjin L Njo3
1Department of Public Health and Primary Care, National eHealth Living Lab (NeLL), Leiden University Medical Centre, Leiden, the Netherlands; Greiner Bio-One, Kremsmünster, Austria.
Background:
Alternative means of blood sampling continue to grow due to the scarcity of phlebotomists and the need for person-centered care. It is crucial to consistently support these alternative blood sampling innovations with scientific evidence to guarantee the quality of care, especially when implementing for instance Lateral Upper-arm Blood Collection (LUBC) for use at home. Knowledge gaps remain in how to quantify imprecision introduced by the collection method and its impact on clinical use.
Methods:
We developed the Six Sigma analysis for alternative Sampling methods, including sampling method-induced imprecision, accuracy and precision, called σ3-methodology (S3). We performed a two-step verification using σ3-methodology for LUBC TAP® II, which included fourteen routine clinical analytes. We used venipuncture as the gold standard.
Results:
The biggest source of imprecision for all analytes was sampling method-induced imprecision, which, for the first time, was quantified, resulting in a varying effect on the clinical usability. TAP® II showed acceptable analytical performance for ALP, bilirubin, HDL-cholesterol, CRP and sodium. Unacceptable analytical performance was found for ALT, AST, cholesterol, creatinine, GGT, HbA1c, potassium, LDH and triglycerides.
Conclusion:
Alternative blood sampling innovations hold promise for advancing diagnostic care, aiming to deliver accessible decentralised sample collection at home that does not require phlebotomist involvement. However, sampling method-induced imprecision should not be overlooked in the performance assessment to guarantee responsible development that will contribute to the success and desirable societal impact of alternative sampling technologies.
Related Concept Videos
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Assessment of blood pressure in brachial artery(two-step method)
Estimating Population Standard Deviation
Uniform Distribution
Two essential properties of this distribution are
Pre-Procedural Guidelines for Assessing Blood Pressure
Sampling Distribution