Related Experiment Video
Updated: May 2, 2026

Dynamic Multiparameter Platelet Function Assessment Using a Capacitive Biosensor
Published on: May 2, 2025
Preparation and characterization of a sulindac sensor based on PVC/TOA-SUL membrane
1Faculty of Chemistry, Maria Curie-Skłodowska University, 5 M. Curie-Skłodowskiej Square, 20-031 Lublin, Poland.
Abstract:
A potentiometric sulindac sensitive sensor based on tetraoctylammonium (Z)-5-fluoro-2-methyl-1-[[p-(methylsulfinyl)phenyl]methylene]-1H-indene-3-acetate (TOA-SUL) was described. The electrode responded with sensitivity of 57.5±1.6mV decade(-1) over the linear range 5×10(-5)-1×10(-2)mol L(-1) at pH6.0-9.0. It had the limit of detection 1.4×10(-5)mol L(-1), a fast response time of 13s and showed clear discrimination of sulindac ions from several inorganic and organic compounds and also amino acids. This electrode did not contain any inner solutions, so it was easy and comfortable to use. The proposed sensor was used to determine sulindac in clear solution and in urine sample solution.