Published on: August 17, 2022
Intraperitoneal fluid overestimates hydration status assessment by bioimpedance spectroscopy
David Arroyo1, Nayara Panizo1, Soraya Abad1
1Nephrology Department, Hospital General Universitario Gregorio Marañón, Madrid, Spain.
Background:
Bioimpedance spectroscopy (BIS) is a valuable tool to assess nutrition and volume status in peritoneal dialysis (PD) patients. However, data about the influence of intraperitoneal fluid on body composition measures are conflicting, and there is no clear consensus about whether the abdomen should be drained before the procedure. We designed a comparison study to detect the influence of intra-abdominal fluid on BIS results.
Methods:
We performed 73 pairs of BIS measurements in 34 stable PD patients, first with the peritoneum filled with a 1.36% glucose dialysate solution and then after the solution was drained. Patients stayed in the supine position for 10 minutes before the BIS procedure, and the electrodes were not moved between measures. Clinical and demographic data were collected, as were analytic parameters of nutrition and volume status.
Results:
Fluid overload is overestimated when BIS is performed with a full abdomen (1.82 ± 1.73 L vs 1.64 ± 1.68 L, p = 0.043). We also found a spurious overestimation in extracellular water (16.40 ± 3.21 L vs 16.24 ± 3.16 L, p < 0.001) and in relative overhydration (8.29% ± 6.96% vs 7.14% ± 6.79%, p = 0.017). No differences in intracellular water or parameters of nutrition were found. We observed negative correlations for the extracellular water overestimation with age (r = -0.245, p = 0.037), serum B-type natriuretic peptide (r = -0.366, p = 0.036), body mass index (r = -0.248, p = 0.035), and lean tissue index (r = -0.427, p = 0.001). The difference in extracellular water correlated only with body mass index (r = -0.259, p = 0.039). We also found that, assessed at 50 KHz, whole-body impedance (-4.52 ± 8.37, p = 0.001) and phase angle (-0.08 ± 0.23 degrees, p = 0.002) were both lower when BIS was performed in patients with a full abdomen.
Conclusions:
Fluid overload is overestimated by BIS when performed in patients with dialysate in the peritoneum. The observed differences are greater in younger, more poorly nourished, or less overhydrated patients. If more precise results are required, we suggest that the abdomen be drained before BIS is performed.
More Related Videos
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
07:44Author Spotlight: Implementation of BIVA for Analyzing Disease Risk Factors in Patients with Low Body Cell Mass
Published on: July 14, 2023
Related Concept Videos
Body Water Content and Fluid Compartments
Composition of Body Fluids
Peritoneal Dialysis I: Introduction and Procedure
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Drug Dosing: Obese Patients
Ascites