Bone conduction hearing sensitivity in normal-hearing subjects: transcutaneous stimulation at BAHA vs BCI position
Sabine Reinfeldt1, Bo Håkansson, Hamidreza Taghavi
1* Department of Signals and Systems, Chalmers University of Technology , Gothenburg , Sweden.
Objective:
Bone conduction (BC) stimulation closer to the cochlea has previously been shown to give higher cochlear promontory acceleration measured by laser Doppler vibrometry (LDV). This study is investigating whether stimulation closer to the cochlea also gives improved hearing sensitivity. Furthermore, the study compares shifts in hearing sensitivity (BC thresholds) and ear-canal sound pressure (ECSP).
Design:
BC hearing thresholds and ECSP have been measured for stimulation at two positions: the existing bone-anchored hearing aid (BAHA) position, and a new bone conduction implant (BCI) position that is located closer to the cochlea.
Study Sample:
The measurements were made on 20 normal-hearing subjects.
Results:
Depending on frequency, the ipsilateral hearing threshold was 3-14 dB better, and the ipsilateral ECSP was 2-12 dB higher for the BCI than for the BAHA position, with no significant differences between threshold and ECSP shifts at group level for most frequencies, and individually only for some subjects.
Conclusions:
It was found that both the objective ECSP and the subjective hearing threshold measurements gave similar improvement as previous LDV measurements for stimulation closer to the cochlea. One exception was that the LDV measurements did not show the improved sensitivity for frequencies below 500 Hz found here.