Anti-Nuclear Antibody Screening Using HEp-2 Cells
Published on: June 23, 2014
A comparison of anti-nuclear antibody quantification using automated enzyme immunoassays and immunofluorescence
Renata Baronaite1, Merete Engelhart2, Troels Mørk Hansen2
1Department of Clinical Biochemistry, Gentofte Hospital, University of Copenhagen, 2900 Hellerup, Denmark ; Department of Rheumatology, Gentofte Hospital, University of Copenhagen, 2900 Hellerup, Denmark.
Abstract:
Anti-nuclear antibodies (ANA) have traditionally been evaluated using indirect fluorescence assays (IFA) with HEp-2 cells. Quantitative immunoassays (EIA) have replaced the use of HEp-2 cells in some laboratories. Here, we evaluated ANA in 400 consecutive and unselected routinely referred patients using IFA and automated EIA techniques. The IFA results generated by two independent laboratories were compared with the EIA results from antibodies against double-stranded DNA (dsDNA), from ANA screening, and from tests of the seven included subantigens. The final IFA and EIA results for 386 unique patients were compared. The majority of the results were the same between the two methods (n = 325, 84%); however, 8% (n = 30) yielded equivocal results (equivocal-negative and equivocal-positive) and 8% (n = 31) yielded divergent results (positive-negative). The results showed fairly good agreement, with Cohen's kappa value of 0.30 (95% confidence interval (CI) = 0.14-0.46), which decreased to 0.23 (95% CI = 0.06-0.40) when the results for dsDNA were omitted. The EIA method was less reliable for assessing nuclear and speckled reactivity patterns, whereas the IFA method presented difficulties detecting dsDNA and Ro activity. The automated EIA method was performed in a similar way to the conventional IFA method using HEp-2 cells; thus, automated EIA may be used as a screening test.
More Related Videos
09:39Quantitative Immunofluorescence Assay to Measure the Variation in Protein Levels at Centrosomes
Published on: December 20, 2014
10:05Simultaneous Distinction of Monospecific and Mixed DFS70 Patterns During ANA Screening with a Novel HEp-2 ELITE/DFS70 Knockout Substrate
Published on: January 17, 2018
Related Concept Videos
Immunofluorescence Microscopy
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...