Imprecision and limit of quantification of the new architect 25-hydroxyvitamin D assay
Marcel J W Janssen1, Frans A L Van der Horst2, Johannes S Kamphuis3
1VieCuri Medical Center, Laboratory of Clinical Chemistry and Hematology, Venlo, The Netherlands. marceljanssen@viecuri.nl
Abstract:
We examined imprecision and limit of quantification (LOQ) of the new Architect 25-hydroxyvitamin D immunoassay in a multicenter setting. Mean intracenter interlot imprecision varied from 22.0 to 5.0%CV in the concentration range of 16 - 92 nmol/L. LOQs were locally determined as 10.8, 11.2, and 13.5 nmol/L. We conclude that the Architect 25(OH)D assay provides a precise and analytically sensitive method for routine measurement of 25(OH)D.