Related Experiment Video
Updated: May 2, 2026

In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
4D UTE flow: a phase-contrast MRI technique for assessment and visualization of stenotic flows
Mo Kadbi1, Mj Negahdar, Jung Won Cha
1Electrical and Computer Engineering Department, University of Louisville, Louisville, Kentucky, USA.
Purpose:
Inaccuracy of conventional four-dimensional (4D) flow MR imaging in the presence of random unsteady and turbulent blood flow distal to a narrowing has been an important challenge. Previous investigations have revealed that shorter echo times (TE) decrease the errors, leading to more accurate flow assessments.
Methods:
In this study, as part of a 4D flow acquisition, an Ultra-Short TE (UTE) method was adopted. UTE works based on a center-out radial k-space trajectory that inherently has a short TE. By employing free induction decay sampling starting from read-out gradient ramp-up, and by combining the refocusing lobe of the slice select gradient with the bipolar flow encoding gradient, TEs of ≈1 msec may be achieved.
Results:
Both steady and pulsatile flow regimes, and in each case a range of Reynolds numbers, were studied in an in-vitro model. Flow assessment at low and medium flow rates demonstrated a good agreement between 4D UTE and conventional 4D flow techniques. However, 4D UTE flow significantly outperformed conventional 4D flow, at high flow rates for both steady and pulsatile flow regimes. Feasibility of the method in one patient with Aortic Stenosis was also demonstrated.
Conclusion:
For both steady and pulsatile high flow rates, the measured flow distal to the stenotic narrowing using conventional 4D flow revealed more than 20% error compared to the ground-truth flow. This error was reduced to less than 5% using the 4D UTE flow technique.
Related Concept Videos
Magnetic Resonance Imaging
Urodynamic Studies: Uroflowmetry
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...