Pulse-pressure-enhancing controller for better physiologic perfusion of rotary blood pumps based on speed modulation
Feng Huang1, Xiaodong Ruan, Xin Fu
1From The State Key Laboratory of Fluid Power Transmission and Control, Zhejiang University, Hangzhou, China.
Abstract:
Sufficient pulsation is important for physiologic perfusion if adequate flow is to be guaranteed. A fuzzy control method for rotary blood pumps using active speed modulation is proposed in this article. It maintains the mean aortic pressure to provide sufficient perfusion while it simultaneously enhances the pulse pressure. The controller uses the indices extracted from the aortic pressure as feedback to determine the amplitude and offset of the rectangular speed modulation waveform, which is synchronous with the cardiac cycle. An additional algorithm is included to prevent regurgitation. The controller is tested both in a baroreflex-cardiovascular model and in a preliminary in vitro experiment. Simulation results demonstrate that the controller is able to increase the pulse pressure to approximately 20 mm Hg and at the same time maintains the mean pressure at 100 mm Hg, when heart failure occurs. It is also quite robust under various physiologic disturbances. Experimental results show that the speed modulation can be implemented in real pumps and that the controller is feasible in practice.
Related Concept Videos
Frequency-Domain Interpretation of PD Control
The proportional control gain, combined with the...
Time-Domain Interpretation of PD Control
Consider the example of control of motor torque. Initially, a positive...
PD Controller: Design
Designing a continuous-data controller requires selecting and linking components like adders and integrators, which are fundamental in Proportional,...
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
PID Controller
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...