Monoenergetic extrapolation of cardiac dual energy CT for artifact reduction
Francesco Secchi1, Carlo Nicola De Cecco2, James Vance Spearman3
1Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA IRCCS Policlinico San Donato, Radiology Unit, San Donato Milanese, Italy francesco.secchi@grupposandonato.it.
Background:
Monoenergetic extrapolation of cardiac dual-energy computed tomography (DECT) could be useful in artifact reduction in clinical practice.
Purpose:
To evaluate the potential of monoenergetic extrapolation of cardiac DECT data for reducing artifacts from metal and high iodine contrast concentration.
Material And Methods:
With IRB approval and in HIPAA compliance, 35 patients (22 men, 61 ± 12 years) underwent cardiac DECT with dual-source CT (100 kVp and 140 kVp). Contrast material injection protocols were adapted to the patient's weight using non-ionic low-osmolar 370 mgI/mL iopromide. Datasets were transferred to a stand-alone workstation and dedicated monoenergetic analysis software was used for postprocessing. Reconstructions with the following five photon energies were generated: 40 keV, 60 keV, 80 keV, 100 keV, and 120 keV. Artifact severity was graded on a 5-point Likert scale (0, massive artifact; 5, absence of artifact). The size of artifact and image noise (expressed as HU) in anatomic structures adjacent to the artifact were measured. Quantitative and subjective image quality was compared using Friedman and Wilcoxon tests.
Results:
We observed artifacts arising from densely concentrated contrast material in the superior vena cava (SVC) in 18 patients, from sternal wires in 14, from bypass clips in eight, and from coronary artery stents in seven. Artifact size in monoenergetic reconstructions from 40 to 120 keV decreased from 21.3 to 19 mm for the SVC (P < 0.001), from 8.4 to 2.6 mm for sternal wires (P < 0.001), from 6.4 to 2.2 mm for bypass clips (P < 0.001), and from 5.9 to 2.7 mm for stents (P < 0.001), respectively. The quality score changed from 0.2 to 3.8 for the SVC (P < 0.001), from 0.1 to 4 for sternal wires (P < 0.001), from 0 to 3.9 for bypass clips (P < 0.001), and from 0 to 3.9 for stents (P < 0.001). Lowest noise in adjacent structures was found at 80 keV for the SVC (39.1 HU), for sternal wires (33.3), for bypass clips (26.9), and for stents (33.9).
Conclusion:
A significant reduction of high-attenuation artifacts can be achieved by use of higher monoenergetic energy levels with cardiac DECT. However, image noise in anatomic structures affected by artifacts is lowest at 80 keV, which suggests an evaluation approach that makes use of multiple energy levels for a complete diagnosis.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...