Validity Test of a New Open-Circuit Indirect Calorimeter
Christine M Ashcraft1, David C Frankenfield2
1Department of Clinical Nutrition, Department of Nursing, Penn State Milton S. Hershey Medical Center, Hershey, Pennsylvania Cashcraft@hmc.psu.edu.
Background:
Indirect calorimetry is an accurate way to measure resting metabolic rate. The Deltatrac Metabolic Monitor is considered a criterion standard but is no longer manufactured. New-generation indirect calorimeters have been introduced, but there are limited published validation data comparing these devices to criterion instruments.
Materials And Methods:
A prospective, observational, N-of-1 trial was conducted to validate a new-generation indirect calorimeter against a gold standard device. This design was chosen to minimize and define the degree of biological variation, thus focusing on variation due to the devices. Measurements of gas exchange using both indirect calorimeters were conducted daily for 10 consecutive days. Another set of measurement pairs was conducted using just the criterion device for 10 days. Ninety-five percent confidence intervals of differences were used to test for bias. Precision was defined as repeat measures with one device falling within 5% of the other at least 90% of the time.
Results:
There were no statistically significant differences between the devices for any measured or calculated parameter. Interdevice differences were no larger than intradevice differences using the criterion instrument. The values obtained from the new device were precise and unbiased compared with the values obtained from the gold standard device.
Conclusion:
The new indirect calorimeter measures gas exchange in a reliable and accurate manner compared with a gold standard device. The two devices are equivalent.
More Related Videos
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
Related Concept Videos
Constant Volume Calorimetry
Constant Pressure Calorimetry
Calorimetry
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...