Noninvasive techniques for blood pressure measurement are not a reliable alternative to direct measurement: a
Sara Ribezzo1, Eleonora Spina2, Stefano Di Bartolomeo3
1School of Nursing, University of Trieste, 34100 Trieste, Italy.
Introduction:
Noninvasive blood pressure (NIBP) monitoring methods are widely used in critically ill patients despite poor evidence of their accuracy. The erroneous interpretations of blood pressure (BP) may lead to clinical errors.
Objectives:
To test the accuracy and reliability of aneroid (ABP) and oscillometric (OBP) devices compared to the invasive BP (IBP) monitoring in an ICU population.
Materials And Methods:
Fifty adult patients (200 comparisons) were included in a randomized crossover trial. BP was recorded simultaneously by IBP and either by ABP or by OBP, taking IBP as gold standard.
Results:
Compared with ABP, IBP systolic values were significantly higher (mean difference ± standard deviation 9.74 ± 13.8; P < 0.0001). Both diastolic (-5.13 ± 7.1; P < 0.0001) and mean (-2.14 ± 7.1; P=0.0033) IBP were instead lower. Compared with OBP, systolic (10.80 ± 14.9; P < 0.0001) and mean (5.36 ± 7.1; P < 0.0001) IBP were higher, while diastolic IBP (-3.62 ± 6.0; P < 0.0001) was lower. Bland-Altman plots showed wide limits of agreement in both NIBP-IBP comparisons.
Conclusions:
BP measurements with different devices produced significantly different results. Since in critically ill patients the importance of BP readings is often crucial, noninvasive techniques cannot be regarded as reliable alternatives to direct measurements.
More Related Videos
07:35Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure