Neonatal/infant validation study of the CAS model 740 noninvasive blood pressure monitor with the Orion/MaxIQ NIBP
Sean M Lang1, John S Giuliano, Christopher L Carroll
1aYale University School of Medicine, New Haven bConnecticut Children's Medical Center, Hartford cUniversity of Connecticut School of Medicine, Farmington, Connecticut, USA.
Objective:
Blood pressure monitoring is an essential vital sign when caring for critically ill children. Invasive monitoring is considered the gold standard, but is not always feasible. The following study compared the CAS model 740 noninvasive blood pressure monitor with the Orion/MaxIQ NIBP module with the reference (invasive arterial measurement). We evaluated the validity of the system using the criteria provided by the Association for the Advancement of Medical Instrumentation.
Results:
We performed paired measurements of 29 critically ill neonates and children. For individual paired comparisons, the mean difference in the systolic blood pressure was 2.42 mmHg (SD ± 6.3). The mean difference in the diastolic blood pressure was -1.29 mmHg (SD ± 5.45). The percentage of readings within 5, 10, and 15 mmHg for systolic blood pressure was 65.6, 86.0, and 96.8%, respectively. The percentage of readings within 5, 10, and 15 mmHg for diastolic blood pressure was 77.7, 93, and 95.5%, respectively.
Conclusion:
The CAS model 740 noninvasive blood pressure monitor with the Orion/MaxIQ NIBP module fulfills the accuracy performance criteria of the Association for the Advancement of Medical Instrumentation. This model may allow for rapid and accurate noninvasive blood pressure monitoring in neonates and children.
Related Concept Videos
Measurement of Blood Pressure
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.