Related Experiment Video
Updated: May 2, 2026

An Intra-Tissue Radiometry Microprobe for Measuring Radiance In Situ in Living Tissue
Published on: June 2, 2023
INTERCAL: long-term inter-comparison experiment for dose rate and spectrometric probes
1Bundesamt für Strahlenschutz (Federal Office for Radiation Protection), Willy-Brandt-Straße 5, 38226 Salzgitter, Germany mbleher@bfs.de.
The Schauinsland inter-calibration facility (INTERCAL) enables long-term comparison of dose rate probes. It investigates probe behavior under environmental conditions using reference probes and additional instrumentation.
Area of Science:
- Environmental Science
- Radiation Dosimetry
Background:
- The Schauinsland inter-calibration facility (INTERCAL) was established for long-term comparison experiments.
- Two reference probes, characterized by European Radiation Dosimetry WG3, were installed at INTERCAL.
Purpose of the Study:
- To enable long-term comparison experiments for 20 different dose rate probes from various networks.
- To investigate the long-term behavior of dose rate probes and spectrometry systems under environmental conditions.
Main Methods:
- Installation of two characterized reference probes at the INTERCAL facility.
- Deployment of additional instrumentation for measuring air activity, nuclide-specific dose rates, and environmental parameters (air pressure, temperature, precipitation, soil moisture).
- Conducting two exposure experiments in April 2009 and May 2012.
Main Results:
- Established a facility for inter-calibration of dose rate probes.
- Collected data on environmental parameters alongside radiation measurements.
- Provided a framework for long-term studies of radiation detection equipment.
Conclusions:
- The INTERCAL facility is suitable for long-term inter-calibration experiments.
- Environmental factors can influence dose rate probe performance.
- The platform supports research into radiation detection under real-world conditions.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Atomic Emission Spectroscopy: Lab
UV–Vis Spectrometers