Parallel imaging performance investigation of an 8-channel common-mode differential-mode (CMDM) planar array for 7T
Xiaoqing Hu1, Xiao Chen1, Xin Liu1
11 Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology of Chinese Academy of Sciences, Shenzhen 518055, China ; 2 Shenzhen Key Laboratory for MRI, Shenzhen 518055, China ; 3 Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA, USA ; 4 UCSF/UC Berkeley Joint Graduate Group in Bioengineering, San Francisco, CA, USA.
Abstract:
An 8-channel planar phased array was proposed based on the common-mode differential-mode (CMDM) structure for ultrahigh field MRI. The parallel imaging performance of the 8-channel CMDM planar array was numerically investigated based on electromagnetic simulations and Cartesian sensitivity encoding (SENSE) reconstruction. The signal-to-noise ratio (SNR) of multichannel images combined using root-sum-of-squares (rSoS) and covariance weighted root-sum-of-squares (Cov-rSoS) at various reduction factors were compared between 8-channel CMDM array and 4-channel CM and DM array. The results of the study indicated the 8-channel CMDM array excelled the 4-channel CM and DM in SNR. The g-factor maps and artifact power were calculated to evaluate parallel imaging performance of the proposed 8-channel CMDM array. The artifact power of 8-channel CMDM array was reduced dramatically compared with the 4-channel CM and DM arrays demonstrating the parallel imaging feasibility of the CMDM array.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging