Quantitative Static and Dynamic Assessment of Balance Control in Stroke Patients
Published on: May 17, 2020
Comparison of the test-retest reliability of the balance computerized adaptive test and a computerized posturography
Chia-Hsin Chen1, Shih-Feng Lin2, Wan-Hui Yu3
1Department of Physical Medicine and Rehabilitation, Kaohsiung Municipal Ta-Tung Hospital, Kaohsiung, Taiwan; Department of Physical Medicine and Rehabilitation, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan; Department of Physical Medicine and Rehabilitation, School of Medicine, Graduate Institute of Neuroscience, College of Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan.
Objective:
To compare the test-retest reliabilities of the scores of the Balance Computerized Adaptive Test (CAT) and the Biodex Balance System in patients with stroke.
Design:
A repeated-measures design (at a 1-wk interval) was used to examine the test-retest reliabilities of the scores of the Balance CAT and the Biodex Balance System.
Setting:
One rehabilitation unit in a local hospital.
Participants:
Patients (N=50) with stroke for more than 6 months and undergoing outpatient rehabilitation completed the Balance CAT and the eyes open (EO)/closed (EC) tests, but only 17 patients finished the Limit of Stability (LOS) test because they were unable to reach all the targets.
Interventions:
Not applicable.
Main Outcome Measures:
The Balance CAT and 2 computerized tests of the Biodex Balance System, namely the EO/EC test and the LOS, were used to evaluate balance function.
Results:
The test-retest reliabilities of the scores of the Balance CAT (Pearson r=.92, minimal detectable change [MDC] percent=12.8%) was excellent. Those of the EO/EC and LOS tests were poor to good (Pearson r=.56-.85, MDC%=50.8%-126.9%).
Conclusions:
The test-retest reliabilities of the scores of the Balance CAT were sufficient for assessing balance function in patients with stroke. Moreover, the test-retest reliabilities of the scores of the Balance CAT, one of the functional balance measures, were superior to those of the Biodex Balance System, 1 type of computerized posturography instrument. Therefore, the Balance CAT may be a more reliable measure for clinicians and researchers to use in assessing the balance function of patients with stroke for more than 6 months.