Imaging and Quantification of the Area of Fast-Moving Microbubbles Using a High-Speed Camera and Image Analysis
Published on: September 5, 2020
Cavitation measurement during sonic and ultrasonic activated irrigation
Ricardo Macedo1, Bram Verhaagen2, David Fernandez Rivas3
1Department of Cariology, Endodontology and Pedodontology, ACTA Academic Center for Dentistry Amsterdam, University of Amsterdam and VU University, Amsterdam, The Netherlands.
Introduction:
The aims of this study were to quantify and to visualize the possible occurrence of transient cavitation (bubble formation and implosion) during sonic and ultrasonic (UAI) activated irrigation.
Methods:
The amount of cavitation generated around several endodontic instruments was measured by sonochemiluminescence dosimetry inside 4 root canal models of human dimensions and varying complexity. Furthermore, the spatial distribution of the sonochemiluminescence in the root canal was visualized with long-exposure photography.
Results:
Instrument oscillation frequency, ultrasonic power, and file taper influenced the occurrence and amount of cavitation. In UAI, cavitation was distributed between the file and the wall extending beyond the file and inside lateral canals/isthmuses. In sonic activated irrigation, no cavitation was detected.
Conclusions:
Cavitation was shown to occur in UAI at clinically relevant ultrasonic power settings in both straight and curved canals but not around sonically oscillating instruments, driven at their highest frequency.