A comparison of phase, amplitude, and velocity binning for cone-beam computed tomographic projection-based motion
Nathan Becker1, Sarah Quirk2, Ian Kay3
1Department of Radiation Physics, Princess Margaret Hospital, Toronto, Ontario, Canada.
Purpose:
We previously developed a motion estimation technique based on direct cone-beam projection analysis. It is able to reconstruct the complete motion trajectory of a radio-opaque marker, including cycle-to-cycle variability, using respiratory binning of the projection images. This paper investigates the use of phase, amplitude, and amplitude-velocity binning in the context of projection-based cone-beam motion estimation (CBME).
Methods And Materials:
We simulated cone-beam computed tomographic scans of 160 tumor trajectories estimated by a CyberKnife Synchrony System (Accuray, Sunnyvale, CA), and reconstructed the complete trajectory with CBME using phase, amplitude, and amplitude-velocity binning of the projection data. Various numbers of respiratory bins, from 1 (no binning) to 100, were used for phase and amplitude binning, while 1 to 100 amplitude bins with 4 velocity bins were used for amplitude-velocity binning. From this large pool of data, we correlated the reconstruction accuracy with bin type, total number of bins, number of breathing cycles per bin, and the position of the bin within the breathing cycle.
Results:
CBME predicted the true motion of the marker with a 3-dimensional (3D) mean root mean square (RMS) error of 0.24 mm for amplitude-velocity binning, 0.31 mm for amplitude binning, and 0.52 mm for phase binning. Reconstruction 3D RMS error increased to over 1 mm when less than 3 breathing cycles contributed to a bin. We found that reconstruction accuracy was optimized when about 20 bins were used. Accuracy also decreased in bins located around the inhale portion of the breath cycle, compared with the mid- and end-exhale positions.
Conclusions:
This study provides a quantitative assessment of phase, amplitude, and amplitude-velocity binning for CBME. A joint binning approach should be used to give both the accuracy of amplitude binning, as well as the robustness of phase binning, in areas of limited motion sampling.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Imaging Studies III: Computed Tomography