Initial performance evaluation of iterative model reconstruction in abdominal computed tomography
Related Experiment Video
Updated: May 1, 2026

Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Shigeru Suzuki1, Takahiro Haruyama, Hisashi Morita
1From the *Department of Diagnostic Radiology, Saitama Red Cross Hospital, Saitama, Japan; †Department of Radiology, Teikyo University School of Medicine, Tokyo, Japan; and ‡Department of Radiology, Saitama Red Cross Hospital, Saitama, Japan.
Objective:
The aim of this study was to present initial evaluation of the performance of the iterative model reconstruction algorithm (IMR) in abdominal computed tomography (CT).
Methods:
Computed tomographic examinations were performed for clinical study of 36 patients and for phantom study. We reconstructed the raw data with 1.0- and 5.0-mm slice thicknesses using filtered back projection (FBP), iDose4, and IMR and evaluated image quality objectively and subjectively.
Results:
For almost all subjective characteristics, the image quality was better using IMR than iDose4. Objective image noise was significantly less using IMR than iDose4 (P < 0.0001). The contrast-noise ratio of both slice thicknesses increased in order from FBP to iDose4 to IMR. The spatial resolution of reconstructed images was almost identical using IMR, FBP, and iDose4.
Conclusions:
The IMR can significantly improve image noise and low-contrast resolution and maintain edge sharpness in abdominal CT images compared with iDose4 or FBP.