Evaluation of on-board kV cone beam computed tomography-based dose calculation with deformable image registration
Yusuke Onozato1, Noriyuki Kadoya1, Yukio Fujita1
1Department of Radiation Oncology, Tohoku University School of Medicine, Sendai, Japan.
Purpose:
The purpose of this study was to estimate the accuracy of the dose calculation of On-Board Imager (Varian, Palo Alto, CA) cone beam computed tomography (CBCT) with deformable image registration (DIR), using the multilevel-threshold (MLT) algorithm and histogram matching (HM) algorithm in pelvic radiation therapy.
Methods And Materials:
One pelvis phantom and 10 patients with prostate cancer treated with intensity modulated radiation therapy were studied. To minimize the effect of organ deformation and different Hounsfield unit values between planning CT (PCT) and CBCT, we modified CBCT (mCBCT) with DIR by using the MLT (mCBCT(MLT)) and HM (mCBCT(HM)) algorithms. To evaluate the accuracy of the dose calculation, we compared dose differences in dosimetric parameters (mean dose [D(mean)], minimum dose [D(min)], and maximum dose [D(max)]) for planning target volume, rectum, and bladder between PCT (reference) and CBCTs or mCBCTs. Furthermore, we investigated the effect of organ deformation compared with DIR and rigid registration (RR). We determined whether dose differences between PCT and mCBCTs were significantly lower than in CBCT by using Student t test.
Results:
For patients, the average dose differences in all dosimetric parameters of CBCT with DIR were smaller than those of CBCT with RR (eg, rectum; 0.54% for DIR vs 1.24% for RR). For the mCBCTs with DIR, the average dose differences in all dosimetric parameters were less than 1.0%.
Conclusions:
We evaluated the accuracy of the dose calculation in CBCT, mCBCT(MLT), and mCBCT(HM) with DIR for 10 patients. The results showed that dose differences in D(mean), D(min), and D(max) in mCBCTs were within 1%, which were significantly better than those in CBCT, especially for the rectum (P<.05). Our results indicate that the mCBCT(MLT) and mCBCT(HM) can be useful for improving the dose calculation for adaptive radiation therapy.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography