Comparison of devices for newborn ventilation in the delivery room
Edgardo Szyld1, Adriana Aguilar2, Gabriel A Musante3
1Research Department, FUNDASAMIN, Fundación para la Salud Materno Infantil, Buenos Aires, Argentina.
Objective:
To evaluate the effectiveness and safety of a T-piece resuscitator compared with a self-inflating bag for providing mask ventilation to newborns at birth.
Study Design:
Newborns at ≥26 weeks gestational age receiving positive-pressure ventilation at birth were included in this multicenter cluster-randomized 2-period crossover trial. Positive-pressure ventilation was provided with either a self-inflating bag (self-inflating bag group) with or without a positive end-expiratory pressure valve or a T-piece with a positive end-expiratory pressure valve (T-piece group). Delivery room management followed American Academy of Pediatrics and International Liaison Committee on Resuscitation guidelines. The primary outcome was the proportion of newborns with heart rate (HR)≥100 bpm at 2 minutes after birth.
Results:
A total of 1027 newborns were included. There was no statistically significant difference in the incidence of HR≥100 bpm at 2 minutes after birth between the T-piece and self-inflating bag groups: 94% (479 of 511) and 90% (466 of 516), respectively (OR, 0.65; 95% CI, 0.41-1.05; P=.08). A total of 86 newborns (17%) in the T-piece group and 134 newborns (26%) in the self-inflating bag group were intubated in the delivery room (OR, 0.58; 95% CI, 0.4-0.8; P=.002). The mean±SD maximum positive inspiratory pressure was 26±2 cm H2O in the T-piece group vs 28±5 cm H2O in the self-inflating bag group (P<.001). Air leaks, use of drugs/chest compressions, mortality, and days on mechanical ventilation did not differ significantly between groups.
Conclusion:
There was no difference between the T-piece resuscitator and a self-inflating bag in achieving an HR of ≥100 bpm at 2 minutes in newborns≥26 weeks gestational age resuscitated at birth. However, use of the T-piece decreased the intubation rate and the maximum pressures applied.
More Related Videos
04:56Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
08:58Development of a Neonatal Piglet Acute Lung Injury Model Recreating the Early Environment of Preterm Infant Lungs
Published on: October 31, 2025
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...