Related Experiment Video
Updated: May 1, 2026

Echocardiographic Assessment Using Subxiphoid-Only Examination for Hypotensive Patients
Published on: April 18, 2025
Difficulties encountered by physicians in interpreting focused echocardiography using a pocket ultrasound machine in
Cyril Charron1, François Templier, Nathalie S Goddet
1aIntensive Care Unit, Section Thorax-Vascular Disease-Abdomen-Metabolism, Assistance Publique-Hôpitaux de Paris, University Hospital Ambroise Paré, Boulogne-Billancourt bSAMU 92, Assistance Publique-Hôpitaux de Paris, University Hospital Raymond Poincaré, Garches cFaculty of Medicine Paris Ile-de-France Ouest, University of Versailles Saint-Quentin en Yvelines, Saint-Quentin en Yvelines, France.
Objective:
Pocket ultrasound devices (PUDs) increase the scope of transthoracic echocardiography. We assessed the ability of emergency physicians (EPs) to obtain and interpret views using PUDs in prehospital emergencies.
Materials And Methods:
Nine EPs underwent a 2-day training program focused on acquisition of four views and on evaluation of left ventricular function, right ventricular size, the inferior vena cava, and detection of pericardial effusion. Then, EPs used a PUD to perform transthoracic echocardiography in patients with shock or acute respiratory failure. The quality and interpretation of views were graded by an expert as not obtained/inadequate, adequate, or optimal. Agreement between the expert and the physicians was evaluated using Cohen's κ test.
Results:
One hundred consecutive exams were evaluated in patients with shock or acute respiratory failure. Parasternal long-axis and short-axis views, and a subcostal view were not obtained or inadequate in 56, 54, and 54 patients, respectively. An apical four-chamber view was not obtained or inadequate in 33 patients. One, two, or three views were graded as adequate or optimal in 86, 65, and 35 patients. Agreement between physicians and experts for left ventricular systolic function, right ventricular size, and pericardial effusion was weak [κ 0.37 (0.17; 0.59), 0.27 (0.023; 0.53), and 0.33 (-0.008; 0.67)]. Agreement for inferior vena cava evaluation was very weak [0.13 (-0.17; 0.43)].
Conclusion:
After a very short training program, echocardiography using a PUD in prehospital emergencies was feasible in half of patients. Acquisition of technical skills is reasonable, but accurate evaluation of cardiac function may require more extensive training.
More Related Videos
06:50Troubleshooting FoCUS Image Acquisition: Patient Positioning, Transducer Manipulation, and Image Optimization
Published on: March 3, 2023
03:40Point-of-Care Ultrasound for Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation Without Left Ventricular Venting
Published on: January 17, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...