Analyzing and modeling rheological behavior of liver fibrosis in rats using shear viscoelastic moduli
Ying Zhu1, Yi Zheng, Yuan-yuan Shen
1National-Regional Key Technology Engineering Laboratory for Medical Ultrasound, Department of Biomedical Engineering, School of Medicine, Shenzhen University, Shenzhen 518160, China; Department of Electrical and Computer Engineering, St. Cloud State University, St. Cloud, MN 56301, USA.
Abstract:
The process of liver fibrosis changes the rheological properties of liver tissue. This study characterizes and compares liver fibrosis stages from F0 to F4 in rats in terms of shear viscoelastic moduli. Here two viscoelastic models, the Zener model and Voigt model, were applied to experimental data of rheometer tests and then values of elasticity and viscosity were estimated for each fibrosis stage. The results demonstrate that moderate fibrosis (≤F2) has a good correlation with liver viscoelasticity. The mean Zener elasticity E1 increases from (0.452±0.094) kPa (F0) to (1.311±0.717) kPa (F2), while the mean Voigt elasticity E increases from (0.618±0.089) kPa (F0) to (1.701±0.844) kPa (F2). The mean Zener viscosity increases from (3.499±0.186) Pa·s (F0) to (4.947±1.811) Pa·s (F2) and the mean Voigt viscosity increases from (3.379±0.316) Pa·s (F0) to (4.625±1.296) Pa·s (F2). Compared with viscosity, the elasticity shows smaller variations at stages F1 and F2 no matter what viscoelastic model is used. Therefore, the estimated elasticity is more effective than viscosity for differentiating the fibrosis stages from F0 to F2.