Evaluation of a textile-based near infrared spectroscopy system in calf muscle oxygenation measurements
Nassim Nasseri1,2, Christoph Zysset3, Lars Büthe3
1Biomedical Optics Research Laboratory, Division of Neonatology, University Hospital of Zürich, Frauenklinikstrasse 10, 8091, Zurich, Switzerland. nassim.nasseri@usz.ch.
Abstract:
We recently introduced a novel textile-based NIRS sensor (TexNIRS). Here, we evaluate TexNIRS in ten subjects (16 legs, age 28.5 ± 2.32 years, adipose tissue thickness (ATT) 4.17 ± 1.71 mm). Three venous occlusions at 50 mmHg were performed on their calf muscle. After 3 min of occlusion, oxy/deoxy hemoglobin concentration ([O₂Hb], [HHb]) changes were 3.71 ± 1.89/1.79 ± 1.08 μM; venous oxygen saturation (SvO₂) was 75 ± 9.7 %, oxygen consumption (VO₂) was 0.02 ± 0.01 mL/100 g/min, hemoglobin flow (HF) was 0.93 ± 0.48 μmol/100 mL/min, and blood flow (BF) was 2.01 ± 1.04 mL/100 mL/min. Our results are in good agreement with the literature, but the TexNIRS enables a much higher level of comfort.