Establishing a normal reference range for thromboelastography in North Indian healthy volunteers
Arulselvi Subramanian1, Venencia Albert, Renu Saxena
1Department of Lab Medicine, JPNATC, New Delhi, India.
Indian Journal of Pathology & Microbiology
|April 18, 2014
Summary
This study established normal thromboelastography (TEG) reference values for the North Indian population, finding manufacturer ranges may not be suitable. Establishing local TEG values is recommended for accurate coagulopathy assessment.
Area of Science:
- Hematology
- Coagulation Diagnostics
Background:
- Thromboelastography (TEG) offers real-time coagulation analysis but lacks established reference values for the Indian population.
- Limited published literature exists on TEG assessment in this demographic.
Purpose of the Study:
- To determine normal TEG reference values in healthy North Indian adults.
- To compare these values with conventional coagulation tests and manufacturer-provided ranges.
Main Methods:
- 200 healthy volunteers (100 male, 100 female) aged 20-50 years were enrolled.
- Thromboelastometry (TEM) was performed using whole blood on a TEM-A analyzer.
- Key TEG parameters (r-time, k-time, α-angle, MA) were analyzed; PT, aPTT, and platelet counts were also measured.
Main Results:
- Established 95% reference ranges: r-time (1.8-14.2 min), k-time (0.7-7.3 min), α-angle (27.3-72.3°), MA (32.1-87.9 mm).
- Significant variations (P < 0.001) were observed between manufacturer and derived reference ranges for all TEG parameters.
- Manufacturer ranges identified 74% of volunteers as having deranged TEG values, suggesting oversensitivity.
Conclusions:
- Manufacturer TEG reference values may not be appropriate for the North Indian population due to ethnic variations.
- TEG is sensitive but may overdiagnose coagulopathy without clinical correlation and cannot replace conventional tests.
- Laboratories should establish their own normal TEG values for their specific populations.