Related Experiment Video
Updated: Apr 30, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Protection from outpatient sudden cardiac death following ICD removal using a wearable cardioverter defibrillator
Tanyanan Tanawuttiwat1, Juan D Garisto, Arturo Salow
1Division of Cardiovascular Disease, Department of Medicine, University of Miami Miller School of Medicine, Miami, Florida.
Background:
An implantable cardioverter defibrillator (ICD) is effective in preventing sudden cardiac death (SCD). Once an ICD is removed and reimplantation is not feasible, a wearable cardioverter defibrillator (WCD) may be an alternative option. We determined the effectiveness of WCD for SCD prevention in patients who were discharged after ICD removal.
Methods:
A retrospective study was conducted on all WCD (LifeVest, ZOLL, Pittsburgh, PA, USA) patients who underwent ICD removal due to cardiac device infections (CDIs) at two referral centers between January 1, 2005 and December 31, 2009. Clinical characteristics, device information, and WCD data were analyzed. Sudden cardiac arrest was defined as all sustained ventricular tachycardia (VT) and ventricular fibrillation occurring within a single 24-hour period.
Results:
Ninety-seven patients (mean age 62.8 ± 13.3, male 80.4%) were included in the study. The median duration of antibiotic use was 14.7 days (interquartile range [IQR] 10-30). The median daily WCD use was 20 hours/day and the median length of use was 21 days (IQR 5-47). A total of three patients were shocked by WCD. Two patients had four episodes of sustained VT, successfully terminated by the WCD. A third patient experienced two inappropriate treatments due to oversensitivity of the signal artifact. Three patients experienced sudden death outside the hospital while not wearing the device. Five patients died while hospitalized.
Conclusion:
WCD can prevent SCD, until ICD reimplantation is feasible in patients who underwent device removals for CDI. However, patient compliance is essential for the effective use of this device.
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy II: Dilated Cardiomyopathy
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Dysrhythmias VI: Management of Dysrhythmias