Osstell Resonance Frequency Measurement Values as a Prognostic Factor in Implant Dentistry
Tim Krafft1, Friedrich Graef2, Matthias Karl3
11 Private practice, Weiden, Germany.
Abstract:
Resonance frequency analysis (RFA) using the Osstell device (Osstell AB, Gothenburg, Sweden) has been advocated for quantifying implant stability on a relative scale of implant stability quotients (ISQ). It was the goal of this prospective clinical study to evaluate whether a certain ISQ level, at the time an implant is placed, correlates with successful osseointegration as some have claimed. Four hundred ninety-five implants (Straumann AG, Basel, Switzerland), varying in length and diameter, were placed in a private practice, strictly adhering to the implant manufacturer's surgical protocol. After placement and after healing periods of 42 days in the mandible and 56 days (implant manufacturer's protocol) in the maxilla, implant stability was measured using RFA. After healing, implants were torqued forward at 35 Ncm and allowed to heal further if the patients felt discomfort. Statistical analysis of the data obtained was based on Welch tests and Kolmogorov-Smirnow tests (level of significance α = 0.05). Results showed that 432 implants were osseointegrated after the predefined healing periods while 8 implants were lost and, in 55 cases, healing was prolonged. Both at insertion (P = .025) and after healing (P < .001), successful implants showed significantly different ISQ values as compared to implant failures or implants with prolonged healing. However, overlapping ISQ distributions at implant insertion demonstrated that there was no correlation among the data that could be used to predict successful osseointegration. Within the limits of this study, the prognostic value of ISQ values appears to be ambiguous.