Development of an In Vitro Ocular Platform to Test Contact Lenses
Published on: April 6, 2016
10.4K
Dynamic contour tonometry over silicone hydrogel contact lens
Andrew K C Lam1, Jimmy S H Tse1
1School of Optometry, The Hong Kong Polytechnic University, kowloon, Hong Kong.
Journal of Optometry
|April 29, 2014
Summary
Dynamic Contour Tonometry (DCT) is feasible over silicone hydrogel contact lenses. Low modulus lenses showed no clinical effect on intraocular pressure (IOP) measurements, improving agreement.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Optometry
Background:
- Accurate intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
- Silicone hydrogel contact lenses are widely used, but their effect on IOP measurements requires investigation.
- Dynamic Contour Tonometry (DCT) offers a non-contact method for IOP and ocular pulse amplitude (OPA) assessment.
Purpose of the Study:
- To compare IOP and OPA measurements using DCT with and without silicone hydrogel contact lenses of varying modulus.
- To evaluate the impact of different silicone hydrogel lens moduli on corneal biomechanics measured by Ocular Response Analyzer (ORA).
- To assess the feasibility and clinical effect of DCT over in-situ silicone hydrogel contact lenses.
Main Methods:
- Forty-seven young subjects underwent IOP, OPA, corneal hysteresis (CH), and corneal resistance factor (CRF) measurements.
- Measurements were taken without lenses and with two brands of silicone hydrogel contact lenses, with random assignment and crossover.
- DCT and ORA were utilized to assess ocular parameters and corneal biomechanics.
Main Results:
- No significant differences in CH and CRF were observed across different contact lens conditions.
- IOP measurements showed greater variability over high modulus silicone hydrogel lenses compared to no lens.
- Improved agreement for IOP measurements was noted over low modulus silicone hydrogel lenses, with OPA agreement within 1.0mmHg.
Conclusions:
- DCT is a feasible method for measuring IOP and OPA over silicone hydrogel contact lenses.
- Low modulus silicone hydrogel contact lenses do not clinically affect DCT measurements.
- The findings support the use of DCT in patients wearing low modulus silicone hydrogel contact lenses.
Keywords:
Amplitud del pulso ocularCorneal hysteresisCorneal resistance factorFactor de resistencia cornealHistéresis cornealIntraocular pressureLens modulusMódulo de la lenteOcular pulse amplitudePresión intraocular