Posterior pole asymmetry analysis with optical coherence tomography
L Kochendörfer1, P Bauer2, J Funk1
1UniversityHospital Zurich, Department of Ophthalmology, Zurich, Switzerland.
Background:
The aim of this study was to evaluate the reproducibility of the posterior pole asymmetry analysis (PPAA) in glaucomatous eyes and healthy controls in comparison to retinal nerve fiber layer measurements (RNFL) with spectral domain optical coherence tomography (SD-OCT).
Patients And Methods:
Prospectively, in a single session by one operator, three measurements were taken of the retinal thickness with the PPAA and three measurements of the RNFL with the eye-tracker function of the Heidelberg Spectralis SD-OCT.
Results:
The mean intraclass correlation coefficient (ICC) and lower confidence intervals (in parentheses) for PPAA were 0.98 (0.96) in both healthy (36 eyes) and glaucomatous eyes (26 eyes). Global coefficients of variation (COVs) were between 0.0 and 1.1% in healthy controls and between 0.0 and 1.9% in glaucoma eyes. Global ICCs for RNFL were 0.98 (0.96) in healthy controls and 0.98 (0.97) in glaucoma eyes and global COVs were between 0.0 and 6.8% in healthy controls and between 0.0 and 3.0% in glaucoma eyes.
Conclusion:
Reproducibility of the PPAA was good and exceeds that of established RNFL measurements. Thus, PPAA is an additional useful tool in management of glaucoma.
More Related Videos
07:44In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
07:08Using Optical Coherence Tomography and Optokinetic Response As Structural and Functional Visual System Readouts in Mice and Rats
Published on: January 10, 2019
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...