Phase-contrast and magnification radiography at diagnostic X-ray energies using a pseudo-microfocus X-ray source
11 Christie Medical Physics and Engineering, The Christie NHS Foundation Trust, Manchester, UK.
Objective:
To investigate the use of conventional diagnostic X-ray tubes for applications in which specialist microfocus sources are normally required.
Methods:
A conventional diagnostic X-ray tube was used in conjunction with a range of apertures to investigate improvements in spatial resolution using a line-pairs test object. Phase-contrast effects were investigated by varying source-to-object and object-to-receptor distances using a 2-French catheter as a clinically realistic test object.
Results:
For magnification radiography using a computed radiography receptor and conventional X-ray tube with a 1-mm nominal focus size, the limiting spatial resolution was improved from 3.55 line-pairs per millimetre, for a conventional contact image, to 5.6 line-pairs per millimetre, for a ×2 magnified view with a 250-µm aperture. For inline phase-contrast radiography, phase contrast enhancement of a 2-French catheter was demonstrated, and the expected trends with variations in source-to-object and object-to-receptor distances were found. Images of a neonatal phantom demonstrated a subtle improvement in visibility of a superimposed 1-French catheter simulating a percutaneously inserted central catheter for no increase in patient radiation dose.
Conclusion:
Spatial resolution improvement and visible phase contrast can be produced in clinically relevant objects using a pseudo-microfocus geometry at X-ray energies in the normal diagnostic range, using conventional diagnostic X-ray tubes and image receptors. The disadvantages of the proposal are the large distances required to produce phase contrast and limitations imposed by the resulting tube loading.
Advances In Knowledge:
It is possible to use conventional diagnostic X-ray equipment in applications that normally require microfocus X-ray sources. This presents some possibilities for clinical applications.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
X-ray Imaging
Overview of Microscopy Techniques
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...