Radiation protection methods for the interventionalist's hands: use of an extension tube
Shaheen Dixon1, Daniel Schick, John Harper
1Department of Interventional Radiology, Princess Alexandra Hospital, 199 Ipswich Road, Woolloongabba, QLD, 4102, Australia, shaheen7noorani@gmail.com.
Purpose:
Cumulative radiation exposure to the hands during certain interventional procedures may be high. It is important to decrease the amount of radiation to the operator due to the possibility of deterministic effects. We performed a pilot study to demonstrate a significant decrease in operator dose when using extension tubing (ET) in combination with shielding and collimation during a simulated percutaneous transhepatic cholangiogram (PTC) procedure.
Methods:
A whole body, anthropomorphic phantom was used to simulate the patient. A Unfors-Xi Survey detector (to measure scatter) supported by a retort stand and trolley was placed in various positions to simulate the position of hands and eyes/thyroid of an interventionalist. Radiation dose was measured simulating left and right-sided PTC punctures with and without a lead shield, and with and without ET.
Results:
Regarding the radiation dose to the hands; the use of an ET reduces dose by 54 % in right-sided PTC punctures without a shield and by 91 % if used in combination with a shield. For left-sided PTC punctures, ET reduces hand dose by 75 %. The use of collimation decreases hand dose by approximately 60 %. The use of shielding reduces dose to the eyes/thyroid by 98 %.
Conclusions:
The dose to the hands can be significantly reduced with the appropriate use of a shield, ET, and tight collimation. The use of a shield is paramount to reduce dose to the eyes/thyroid. It is important for interventionalists to adhere to radiation protective practice considering the potential deterministic effects during a lifelong career.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...