Comparative evaluation of three different methods for HbA1c measurement with High-performance liquid chromatography
Azadeh Karami1, Azar Baradaran1
1Department of Pathology, Isfahan University of Medial Science, Iran.
Background:
The global prevalence of diabetes mellitus is increasing rapidly. Measurement of glycated hemoglobin, predominantly HbA1c, is fundamental to the management of patients with diabetes. HbA1c is used to monitor long-term glycemic control, adjust therapy, assess the quality of diabetes care and predict the risk for the development of complications. While HbA1c is the standard method for long-term glycemic control in diabetic patients, there are different methods for measurement of HbA1c and all laboratories do not use the reference method (high-performance liquid chromatography [HPLC]). The objective of this study is comparison of three different methods with HPLC to find out which method has an acceptable concordance and correlation with the reference method.
Materials And Methods:
Fifty-eight diabetic patients were assessed in this study. The blood sample of each patient was checked with Diazyme (enzymatic assay), Nycocard (boronate-affinity binding) and Biosystem (micro column chromatography). The values of HbA1c of each method were compared with the Knauer-HPLC results.
Results:
THE MEANS OF THE DIFFERENTIAL VALUES BETWEEN EACH METHOD AND HPLC IN THE ANOVA TEST ARE AS FOLLOWS: M = 1.8, SD = 1.09 for Nycocard-HPLC; M = 1.5, SD = 1.08 for biosystem-HPLC; M = 1.3, SD = 1.2 for Diazyme-HPLC. Pearson's correlation coefficient between HPLC and Nycocard; 0.76, HPLC and Diazyme; 0.75 and between HPLC and Biosystem was 0.68. Linear regression parameters for each method with HPLC were also determined.
Conclusion:
Diazyme had a better performance and showed a greater concordance with HPLC among others, although it was not an ideal alternative for HPLC.
More Related Videos
07:24A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
12:02Automated Modular High Throughput Exopolysaccharide Screening Platform Coupled with Highly Sensitive Carbohydrate Fingerprint Analysis
Published on: April 11, 2016
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Elution Process