Related Experiment Video
Updated: Apr 30, 2026

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
[Contact analysis between artificial humeral head and glenoid fossa during humeral external rotation]
Abstract:
We developed a three-dimensional finite element model of the shoulder glenohumeral joint after shoulder arthroplasty including humerus shaft, scapular, scapular cartilage and eight muscles, while each of the muscles was simulated with 50 spring elements. To reduce the element number and improve the analytical precision, we used mixed tetrahedral and hexahedral elements in the model. We then used the model to calculate the biomechanics of the shoulder glenohumeral joint after hemiarthroplasty during humeral external rotation. Results showed that the maximum joint reaction force was 374.72 N and the maximum contact stress was 6. 573 MPa together with the contact areas at 40 degrees external rotation. These might be one of the reasons for prosthetic disarticulation, and would provide theoretical bases to prosthetic design.
Related Concept Videos
Bones of the Upper Limb: Humerus
Muscles that Move the Arm
The pectoralis major has two origins. Its clavicular head originates on the medial half of the clavicle. In contrast, the sternocostal head originates on the costal cartilages of ribs 1-6, the sternum, and the aponeurosis of the external oblique of the...
Bones of the Upper Limb: Ulna