Model-based iterative reconstruction compared to adaptive statistical iterative reconstruction and filtered
Eric W Olcott1, Lewis K Shin1, Graham Sommer2
1Department of Radiology, Stanford University School of Medicine, 300 Pasteur Drive, H1307, Stanford, CA 94305-5105; Veterans Affairs Palo Alto Health Care System, Palo Alto, CA.
Rationale And Objectives:
To prospectively evaluate the perceived image quality of model-based iterative reconstruction (MBIR) compared to adaptive statistical iterative reconstruction (ASIR) and filtered back-projection (FBP) in computed tomography (CT) of the kidneys and retroperitoneum.
Materials And Methods:
With investigational review board and Health Insurance Portability and Accountability Act compliance, 17 adults underwent 31 contrast-enhanced CT acquisitions at constant tube potential and current (range 30-300 mA). Each was reconstructed with MBIR, ASIR (50%), and FBP. Four reviewers scored each reconstruction's perceived image quality overall and the perceived image quality of seven imaging features that were selected by the authors as being relevant to imaging in the region and pertinent to the evaluation of high-quality diagnostic CT.
Results:
MBIR perceived image quality scored superior to ASIR and FBP both overall (P < .001) and for observations of the retroperitoneal fascia (99.2%), corticomedullary differentiation (94.4%), renal hilar structures (96.8%), focal renal lesions (92.5%), and mitigation of streak artifact (100.0%; all, P < .001). MBIR achieved diagnostic overall perceived image quality with approximately half the radiation dose required by ASIR and FBP. The noise curve of MBIR was significantly lower and flatter (P < .001).
Conclusions:
Compared to ASIR and FBP, MBIR provides superior perceived image quality, both overall and for several specific imaging features, across a broad range of tube current levels, and requires approximately half the radiation dose to achieve diagnostic overall perceived image quality. Accordingly, MBIR should enable CT scanning with improved perceived image quality and/or reduced radiation exposure.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies V: Intravenous Urography and Retrograde Pyelography