Insensible water loss through adult extracorporeal membrane oxygenation circuit: an in vitro study
Chang Li Li1, Tam Oi Yan, Kwan Ming Chit Arthur
1From the Department of Intensive Care, Pamela Youde Nethersole Eastern Hospital, Chai Wan, Hong Kong.
Abstract:
Patients on extracorporeal membrane oxygenation (ECMO) are critically ill, and fluid balance need to be managed as accurately as possible. Previous studies have focused on insensible water loss through neonatal ECMO circuit and showed that water loss through the ECMO circuit was correlated with the sweep-gas flow rates. Current study is the first study focusing on insensible water loss through adult ECMO circuit. An in vitro extracorporeal circuit consisting of Jostra Quadrox D membrane oxygenator and centrifugal pump was primed with normal saline. The amount of water loss through the ECMO circuit was found to be linearly correlated with the sweep-gas flow rate and fluid temperature. For every liter per minute of sweep-gas flow at 37°C, 0.046 ml/min of water will be lost, and for every change of fluid temperature by 1°C, water content loss will be changed by 0.0026 ml/min by multiple linear regression (R = 0.996). The average daily water loss for every liter per minute of sweep-gas flow at 33, 34, 35, 36, 37, 38, and 39°C were 51.3, 55, 58.8, 62.5, 66.2, 70.0, and 73.7 ml/day, respectively.
Related Concept Videos
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis
Regulation of Water Output
Peritoneal Dialysis I: Introduction and Procedure
Hemodialysis II: Procedure and Complications
What Are Osmoregulation and Excretion?