Related Experiment Video
Updated: Apr 30, 2026

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
Reducing the impact of wind noise on cochlear implant processors with two microphones
Kostas Kokkinakis1, Casey Cox2
1Speech-Language-Hearing: Sciences and Disorders, University of Kansas, 1000 Sunnyside Avenue, Lawrence, Kansas 66045 kokkinak@ku.edu.
Abstract:
Behind-the-ear (BTE) processors of cochlear implant (CI) devices offer little to almost no protection from wind noise in most incidence angles. To assess speech intelligibility, eight CI recipients were tested in 3 and 9 m/s wind. Results indicated that speech intelligibility decreased substantially when the wind velocity, and in turn the wind sound pressure level, increased. A two-microphone wind noise suppression strategy was developed. Scores obtained with this strategy indicated substantial gains in speech intelligibility over other conventional noise reduction strategies tested.

