Doppler blood pressure measurement in pigs during anaesthesia
Gabrielle C Musk1, Renata S Costa1, Jonathan Tuke2
1School of Veterinary and Life Sciences, Murdoch University, Australia.
Abstract:
To determine the accuracy of Doppler blood pressure measurement in anaesthetised pigs eight large white pigs (Sus scrofa) were anaesthetised with zolazepam and tiletamine as Zoletil 100 (4 mg/kg) and xylazine (2 mg/kg) by intramuscular injection followed by isoflurane in 100% oxygen. Blood pressure measurements were recorded using a Doppler probe on the radial artery and a catheter in the femoral artery connected to a transducer. The sample mean of the Doppler measurements were compared with the sample mean of the invasive systolic arterial blood pressure while accounting for the heart rate, end-tidal CO2 and temperature. The predicted error of the Doppler was greatest when pigs were hypotensive and normothermic: 11 mmHg (CI: 6-15). Doppler measurements should be interpreted with caution in anaesthetised pigs that are hypotensive and normothermic.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: