Testing of the VocaLog Vocal Monitor
1Department of Hearing and Speech, University of Kansas Medical Center, Kansas City, Kansas.
Objective:
To elucidate capabilities and limitations of the VocaLog, a device marketed to log-calibrated dB sound pressure level (SPL).
Study Design:
The study design varied depending on the experiment. All were prospective. Some were case series, and others were cohort studies without controls.
Method:
Experiments were conducted to determine (1) whether the VocaLog logged phonatory activity and silence when it should, (2) if nonphonatory activities were detected, (3) correlation of VocaLog dB values to an external sound level meter (SLM), and (4) accuracy of phonation time (PT) and speaking time (ST) estimates from the VocaLog.
Results:
Silence and phonatory activity were logged as such nearly 100% of the time. Nonphonatory activities were sometimes detected as dB values, including coughs, throat clear, belching, and swallows. The dB values from the VocaLog were strongly correlated with dB SPL from an external SLM. When on the neck, the device rarely picked up external sounds when the external noise was between 85 and 103 dB SPL. The VocaLog gave a reasonable estimate of ST but overestimated PT.
Conclusions:
Overall, the VocaLog holds promise as means of indexing vocal loudness via calibrated dB SPL levels. However, some nonphonatory activity is also likely to be logged. The device provides a reasonable estimate of ST, but not PT.
Related Concept Videos
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Auditory Perception
Holter Monitor: 24-Hour Monitoring
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: